adia control solution “low” for performing a control solution test with the adia blood glucose meter: The results for this control solution should be in the range of 35 - 65. When you use your adia blood glucose meter for the first time, you do not need a control solution. The device is calibrated by default. What do you need the control solution for? Perform a control solution test before using a new container of test strips if you suspect that the blood glucose meter or test strips are not working accurately, if the reading does not match how you feel, or if you think the results are inaccurate. You should also perform a control solution test if you have dropped the blood glucose meter or after disinfecting it. For quality control purposes in healthcare facilities. Which control solution should I use? Use the control solution (high, medium, low) that is close to your measured blood glucose levels. High: 225–345, Medium: 80–120, Low: 35–65. The control solution test should also fall within these control solution ranges. The test with the control solution serves to check the measuring device and the test strips.